PLW 6011 [DP SC CL H14] - Laboratory washer with 2 wash levels in SlimLine design
   

PLW 6011 [DP SC CL H14]
Laboratory washer with 2 wash levels in SlimLine design with steam condenser and chamber lighting.

Price:  
:  
PLW6011 [DP SC CL H14]

More product information

Laboratory washer with 2 wash levels in SlimLine design with steam condenser and chamber lighting.


  • Performance/cycle 84 narrow-necked glasses or 121 pipettes – 168 l usable capacity

    Further loading options

    Alternatively, 312 test tubes or 84 laboratory flasks can be reprocessed per cycle.
  • Ergonomic loading and unloading – telescopic runners
  • Flexible – 3 different level positions

    SmartLoad

    Wash and dry with hot air on two separate wash levels, can be used flexibly in three positions.
  • Have everything in view thanks to the full-glass door

    Full-glass door

    In combination with the chamber lighting, the glass door allows a visual check of the current process.

EAN: 8051160653447 / Article no. 62601102GB / Mat. no. 10961260

Product details - PLW 6011 [DP SC CL H14]

Construction type and version 
Construction type
Freestanding, 1 glazed hinged door

Full-glass door

In combination with the chamber lighting, the glass door allows a visual check of the current process.
Type of heating
Electricity
Front/back
Stainless Steel
Side panels
Stainless steel
Slimline construction type

SlimLine design

The slim design, just 650 mm wide, fits in the smallest of spaces while ensuring a high capacity.
Number of wash levels
2
Loading and unloading with telescopic runners

Telescopic runners

The telescopic runners in the wash chamber make it even easier to load and unload items.
Electric door lock
Buzzer, acoustic signal at end of programme
Service-friendly design

Easy-to-service construction

Short service times thanks to maintenance-optimised appliance design.
Application 
Suitable for laboratories

Perfect for laboratories

These laboratory washers meet the requirements of a laboratory particularly well. Thanks to their powerful wash technology and variety of programmes including flexible loading variations to accommodate all kinds of laboratory glassware, they facilitate fast and analytically pure reprocessing.
Capacity 
Narrow-necked glassware per cycle [number]
84
Test tubes per cycle [number]
312
Laboratory flasks per cycle [number]
84
Pipettes per cycle [number]
121
Performance data 
Circulation pump, Qmax in l/min
700
Wash cabinet usable capacity in l
168
Control system 
Control type
MP.CON 3
Programme selection
Touch on Glass
Programmability
Freely programmable
Programmes [number]
18
Free programme positions [number]
20
Automatic programme restart

Programme failure protection

Time-saving and safe: The programme failure protection facilitates the continuation of the cleaning and disinfection process after a short interruption, such as a power cut. If the success of the reprocessing process should be at risk, the appliance issues a corresponding warning message.
Time left display
Programme sequence display
Selectable display languages

Adjustable display language

No misunderstandings: The user can select his/her language of choice from a number of different options. All display notifications appear in this language, so that the highest level of operating safety is ensured.
Standard electrical connection 
Electrical connection
3N AC 380-400V 50HZ
Heater rating in kW
7,5
Total rated load in kW
8,25
Fuse rating in A
16
Length of mains connection cable in m
2
Water connection/drainage 
Cold water [number]
1
Hot water [number]

Efficient hot water connection

Connection to the warm water supply saves a lot of time and energy.
1
DI water [number]

DI water connection

Thanks to the connection for fully demineralised water (DI water), these washer-disinfectors and thermal disinfectors facilitate optimum reprocessing results. Carrying out the final rinse with DI water also protects valuable wash items from deposits and corrosion.
1
Required flow pressure in kPa
100-800
Maximum water hardness (cold water/hot water) in mmol/l
0,7
Drain pump [DN]
22
Drying unit 
Drying unit heating type
Electricity
Air throughput in m³/h
150
Time setting in 1-minute increments
0-120
HEPA filter class
H14
HEPA filter filtration efficiency (DIN EN 1822) in %
99,995
Electric heater rating in kW
4
Dimensions and weight 
External dimensions, net height in mm
1715
External dimensions, net width in mm
650
External dimensions, net depth in mm
687
External dimensions, gross height in mm

Note 

Including packaging 
1860
External dimensions, gross width in mm

Note 

Including packaging 
945
External dimensions, gross depth in mm

Note 

Including packaging 
840
Wash cabinet, height in mm
586
Wash cabinet, width in mm
555
Wash cabinet, depth in mm
580
Rack height above floor in mm
850
Net weight in kg
192
Gross weight in kg

Note 

Including packaging 
200
Maximum floor load in N/m²

Maximum floor bearing capacity

Maximum floor bearing capacity states the maximum weight (N) that can be placed on the supporting surface on which the machine is installed. This maximum load bearing capacity is the result of adding up the weight of the machine, the load and the water in it, as well as the dynamic forces that occur during operation.
2500
Emissions data 
Sound pressure LpA in cleaning and drying according to DIN EN ISO 11201

Emissions - sound pressure at workplace

The emission sound pressure level depends on the programme and is a value given to the noise created by the machine. Miele Professional appliances have very low noise levels.
<lt/>70 dB(A) re 20 µPa
Heat dissipation into the room in MJ/h

Heat output into the room

During operation warm air is vented into the surrounding room. The released amount of heat/time depends on the set temperatures and the intensity of use. The stated maximum value can be used to control the temperature of the area or to calculate an increase in room temperature.
1,58
Programmes 
Thermal disinfection
Laboratory glassware reprocessing
Plastics
Features 
2 integrated hose dispenser pumps for liquid media

Integrated dispenser pump

Integrated dispenser pumps facilitate the quick and precise dosing of liquid media.
Steam condenser

Steam condenser

Effective moisture removal from exhaust air.
Volume flow control
Interface for process documentation

Process documentation

Safety through the clear documentation of all the reprocessing steps.
Multi-component filter system
Load carrier – direct coupling

Load carriers - direct connection

The direct connection guarantees the precise feed-in of water and drying air (in appliances with hot air drying) and the economical use of resources.
Integrated hot air drying

Integrated hot-air drying

All reprocessing items are dried efficiently using hot air filtered with a HEPA filter.
Setting options for process chemicals (3 x 5 l)

Setting options for media canisters

Neat and tidy: The storage unit for liquid media.
Wash cabinet made from high-quality stainless steel (1.4404/316L)

Wash cabinet grade 316L 

The wash cabinet is made of high-quality stainless steel 1.4404 (AISI 316L).
Plinth
Chamber lighting
Connection options 
RS 232 serial interface

RS 232 serial interface

The serial interface facilitates the exchange of data with other appliances, e.g. for process documentation.
Standards, test certificates and symbols 
CE
EN 61010-2-040
RoHS guideline 2011/65/EU
Machinery Directive 2006/42/EC
Illustration(s) as an example, for information